12 Using the Battery Pack
ARGUS 141 127
12 Using the Battery Pack
Changing the battery pack
Switch the ARGUS off and disconnect the plug-in power supply. Afterwards, loosen the
thumbscrew to release the battery pack.
Battery pack handling
The ARGUS may only be operated with the included battery pack. Connecting any
other voltage supply to the battery pack contacts will damage the ARGUS.
Automatic recharging of the battery pack when the ARGUS is switched off
The ARGUS automatically recharges the battery pack, if the ARGUS is connected to the
plug-in power supply and is switched off and the battery pack voltage is too low. While
charging, the ARGUS displays the message “Charge battery”. If you press and hold the
power switch, the ARGUS will switch off before the battery pack are recharged. The
ARGUS remains on after fully recharging the battery pack.
- The supplied battery pack may only be charged in the ARGUS.
- Do not use the supplied battery pack in other devices.
- Recharge the battery pack fully at least once a month
(even if the ARGUS is not used for a longer period of time).
- If the lithium-ion battery pack is stored, it should first be charged to between 40 and
60 % of its capacity. If the lithium-ion battery pack is stored for a longer period of time,
it should be recharged to this level every six months.
- Please read the extensive notes on safety and the transport of the lithium-ion battery
pack found in the section "Safety Instructions“.